| General specifications | |
|---|---|
| Display Screen | 6“ TFT LCD, 1024x768 pixels
60 Hz update rate |
| Display dimensions | 99x130mm |
| True sampling rate | 200MHz, 12-bit |
| Dimension (W x H x D) | 224 mm x 188 mm x 34 mm |
| Weight | 1.3 kg with battery |
| Battery | Built-in Li-Ion battery 3.6V, 16Ah |
| Battery operating time | Up to 10 hours |
| Languages | Selectable, user-defined custom language |
| External Power Supply for Adaptor | AC 80V~240V 50Hz/60Hz |
| Data Memory | 4 " 16GB (up to 40000 A-Scans) |
| Warranty: | 2 years, optional 3 years |
| Environment Tests | |
| Operation Temperature | -20°C~50°C |
| Storage Environment | -40°C~70°C |
| Model type: | DIO 1000 SFE |
|---|---|
| USB Connector | One " Host USB |
| Encoder Connector | A, B " pulses, TTL 5V, Start |
| Ethernet Connector | External (optional on USB) |
| High Speed Parallel and TTL Port (optional) | Alarm outputs, trigger in/out control |
| Analog Output | Presence of echo in Gate, Thickness, Position of Echo flank in Gate, EchoStart + AUX |
| A/D Sampling Frequency | 200 MHz |
| Probe connector | Mini Lemo® |
| Scan Type | B-scan Thickness / RGB amplitude |
| Focal Law quantity | N/A |
| Scan Angle Range | N/A |
| Scan Setup | N/A |
| Range | 0~29000mm for PRF 100Hz, (Steel longitudinal wave) |
| Material Velocity | 1~19999 m/s |
| Display Delay | -25~29000mm, (Steel longitudinal wave) |
| Auto Transducer Calibration | Zero offset and velocity |
| Units | mm, inch, µs |
| Refracted Angle | Fixed settings of 0°, 30°, 45°, 60°, 70°; variable 10°~90° in 0.1° steps for calculations |
| Test Modes: | Pulse Echo, Dual, Through-Transmission, EMAT |
| Probe Delay Range | -10µs~4800µs with 5ns resolution |
| Gate Monitors | Four Gates: Floating gate, Interface gate, Measuring gate, Back-wall attenuator |
| Alarms | Selectable threshold positive/negative or minimum depth modes |
| Cursors | N/A |
| Display Mode | A, B-Scan, TOFD |
| Auto Gate | Thickness |
| JIS, DAC | 20 points, plus 4 sub-curves |
| TCG, DGS | 20 points |
| Spot weld (optional) | Heuristic evaluation |
| Colour maps | RGB |
| Test Point Selection | Peak, Flank |
| Measurement method | Four Gates: Each can measure echo amplitude, position, alarm
Gate 1: Sound Path, Depth, Surface Distance (curved surface) Gate 2: Thickness Gate 3: Back-wall attenuator Gate 4: Interface Skip diagram for Angle probes |
| Gate Start | Full Range |
| Gate Width | Full Range |
| Gate Trash | 2~98% |
| Pulser | Unipolar Square Wave, Burst |
| Repetition Frequency | 10Hz~20kHz, peak envelope of A-scan display |
| Energy | 18~189V |
| Pulse Width | 15~5000ns |
| Damping | 50, 57, 180 and 1000Ω |
| Pulser Focusing | N/A |
| Pulser Delay | N/A |
| Gain | 0~111dB with 0.1/0.5/1.0/6dB step |
| Digital Gain | -6~36dB used for TCG |
| Bandwith | 0,5~30MHz at -3dB |
| Rectification | Pos HW, Neg HW, FW and RF |
| Amplitude measurement | 0~150% FSH |
| Filter | 1,2, 2.25~2.5, 4, 5,10MHz BP; digital LP 6 to 50MHz |
| Reject | 0~70% |
| Receiver Delay | N/A |