| General specifications | |
|---|---|
| Display Screen | 6“ TFT LCD, 1024x768 pixels 60 Hz update rate |
| Display dimensions | 99x130mm |
| True sampling rate | 200MHz, 12-bit |
| Dimension (W x H x D) | 224 mm x 188 mm x 34 mm |
| Weight | 1.3 kg with battery |
| Battery | Built-in Li-Ion battery 3.6V, 16Ah |
| Battery operating time | Up to 10 hours |
| Languages | Selectable, user-defined custom language |
| External Power Supply for Adaptor | AC 80V~240V 50Hz/60Hz |
| Data Memory | 4 " 16GB (up to 40000 A-Scans) |
| Warranty: | 2 years, optional 3 years |
| Environment Tests | |
|---|---|
| Operation Temperature | -20°C~50°C |
| Storage Environment | -40°C~70°C |
| DIO 1000 LF Low Frequency | |
|---|---|
| Input / Output | |
| USB Connector | One " Host USB |
| Encoder Connector | A, B " pulses, TTL 5V, Start |
| Ethernet Connector | External (optional on USB) |
| High Speed Parallel and TTL Port (optional) | Alarm outputs, trigger in/out control |
| Analog Output | AOUT1: Position of Echo flank in Gate 1 or EchoStart + AUX in Gate 4 AOUT2: Presence of Echo in Gate 1 or Echo Amplitude or Thickness |
| System | |
| A/D Sampling Frequency | 200 MHz |
| Probe connector | Standard Lemo® |
| Number of Channel | 1 channel |
| Scan | |
| Scan Type | B-scan Thickness / R-G-B amplitude |
| Focal Law quantity | N/A |
| Scan Setup | B-scan Thickness Max / Min / Nominal, R-G-B palette modified by Gate1 level |
| Calibration | |
| Range | 0~29000mm for PRF 100Hz (Steel longitudinal wave) |
| Resolution | 5ns for all echo position measurements and timebase settings |
| Material Velocity | 1~19999 m/s |
| Display Delay | -25~29000mm (Steel longitudinal wave) |
| Auto Transducer Calibration | Zero offset and velocity |
| Units | mm, inch, µs |
| Refracted Angle | Fixed settings of 0°, 30°, 45°, 60°, 70°; variable 10°~90° in 0.1° steps for calculations, Skip diagram |
| Test Modes | Dual (Through-Transmission), Dual+EXTamplifier |
| Probe Delay Range | -10µs~4800µs |
| Gates | |
| Gate Monitors | Four Gates: Floating gate, Interface gate, Measuring gate, Back-wall attenuator |
| Alarms | Selectable threshold positive/negative |
| Cursors | N/A |
| Measurement | |
| Display Mode | A, A+B-scan, A+TOFD, A+Beam shape |
| Peak hold | Freezes Echo envelope |
| Auto Gate | Thickness |
| DAC, JIS-DAC | 20 points, linear or dB-linear interpolation, 4 sub-curves (AUX lines) |
| DGS (Distance Gain Size) | BW, SDH, DSR calculation of curve, ERS (Equivalent Reflector Size) |
| TCG (Time Corrected Gain) | for DAC/DGS |
| Attenuation compensation | up to 400.0 dB/m in 0.1 dB/m steps |
| AWS | Curve, value+AWS class |
| API | 5UE Evaluation mode Yes |
| Spot | weld (optional) N/A |
| Curved Surface Correction | Inside, ouside |
| Colour maps | R-G-B |
| Test Point Selection | Peak, Peak+Flank, Peak+Flank+FirstPeak |
| Measurement method | Four Gates: Each can measure echo amplitude, position, alarm. Gate 1: Sound Path, Depth, Surface Distance (curved surface). Gate 2: Thickness. Gate 3: Back-wall attenuator. Gate 4: Interface. Skip diagram for Angle probes |
| Gate Start | Full Range |
| Gate Width | Full Range |
| Gate Trash | 2~98% |
| Pulse Transmitter | |
| Pulser | Unipolar Square Wave, Burst |
| Repetition Frequency | 10Hz~20kHz, peak envelope of A-scan display |
| Energy | 75~275V |
| Pulse Width | 15ns~50µs (10kHz~33MHz) |
| Damping | 180 and 1000Ω |
| Pulser Focusing | N/A |
| Pulser Delay | N/A |
| Receiver | |
| Gain | 0~111dB with 0.1/0.5/1.0/6dB step |
| Digital Gain | -6~36dB used for TCG |
| Bandwith | 10kHz~10MHz at -3dB |
| Rectification | Pos HW, Neg HW, FW and RF |
| Amplitude measurement | 0~150_ FSH |
| Filter | 50k,120k,200k,400k,500k, 10k~200k BP; 10k/100kHz HP; digital LP 6Mhz |
| Averaging | 2 ~ 64x applied on limited range |
| Reject | 0~70% |
| Receiver Delay | N/A |